| Article Index |
|
Page 1
|
|
Page 2
|
Page 2 of 2
The SBS Aluminium Raingauge (ARG)
 Unique aerodynamic profile
 Choice of 0.1 or 0.2mm tips
 Optional internal heaters
 Increased rainfall catch and measurement accuracy
 Conforms to planned British Standard for Tipping Bucket Raingauges
The SBS raingauge is the result of over 3 years extensive fieldwork into determining the optimum design for accurate rainfall measurement in all climatic conditions, irrespective of wind speed and type of precipitation (e.g. drizzle or heavy, convective or tropical rain). The unique design and aerodynamic shape of the SBS raingauge increases measurement accuracy by capturing more rainfall than traditionally shaped cylindrical gauges, minimising outsplash and reducing rainfall losses caused by evaporation.
The funnel of the gauge is made from 2.5mm thick powder coated aluminium while the base is manufactured in LM6 marine grade aluminium. The gauge can therefore be installed in extremely hostile conditions. A version with internal heaters is also available for cold climate installations.
The gauge is compatible with all modern loggers and event recorders including our VF and Hawk series data loggers with optional telemetry. Two reed switches are fitted as standard giving the option of monitoring the gauge on an additional channel if required or alternatively for use as a control output.
OPTIONS AVAILABLE:
SBS500 0.2mm tip gauge with 500sq.cm.collector
SBS500H 0.2mm tip gauge with 500sq.cm.collector and heater
SBS1000 0.1mm tip gauge with 1000sq.cm.collector
SBS1000H 0.1mm tip gauge with 1000sq.cm.collector and heater
Specification:
Operating Temperature
 -30deg C to +55deg C
|
Funnel Rim Height
 440mm |
Reed Contact Rating
 50 watts (dc resistive)
|
Output
 Contact Closure |
Reed Supply Voltage
 100V dc maximum
|
Weight
 6 kgs |
Reed Supply Current
 1A maximum
|
Internal bubble level for accurate installation |
Typical Accuracy
 98%+ at 20mm per hour
 96%+ at 50mm per hour
 95%+ at 120mm per hour
 0.1mm or better at 5mm per hour |
Heated Version
 12V supply - heaters off - 12mA typical
- heaters on - 2.2A typical
 24V supply - heaters off - 15mA typical
- heaters on - 1.1A typical
Heaters activated by internal thermostat when temperature in enclosure falls to 1deg C and de-activated when temperature rises to 4deg C |
|